Safety Evaluation of 4 Red Light Therapy Devices for Myopia

JAMA Ophthalmology 10.1001/jamaophthalmol.2025.5660

February 05, 2026 at 11:00 AM EST

Read the full article

Are commercially available red light therapy devices for myopia considered safe under American National Standards Institute (ANSI) ophthalmic instrument standards?In this quality improvement study of 4 instruments, the widely used EyeRising and the Sky-n1201 reached ANSI safety limits for exposure times below the recommended 180-second treatment duration, whereas the tested light-emitting diode–based device remained within limits for extended exposures.These findings suggest that laser-based red light devices exceed established safety limits and may pose ocular safety risks when used as recommended, highlighting the need for independent validation to balance clinical effectiveness with retinal safety before widespread pediatric use.

Corresponding Author: Lisa A. Ostrin, OD, PhD, University of Houston College of Optometry, 4401 Martin Luther King Blvd, Houston, TX 77204 (lostrin@central.uh.edu).

Link to the article in your story

We encourage you to link out to this article in your story using the link below. It includes an access token that will give free access to the article for your readers up to one year after publication. (The link will be live after the article publishes and embargo is lifted.)

Please see the article for additional information, including full author list, author contributions and affiliations, conflict of interest and financial disclosures, and funding and support.